Abx Micros 60 Hematology Analyzer User Manual

  
  • Hematology

A comparison of the granulocyte count from the ABX Micros ES60 and the neutrophil count from a 5-part differential analyser Author: A. ABX Micros 60 6 - RAB042FEN - User Manual - ABX Micros 60 Once the Identification has been entered, a message will appear on the LCD display stating «Press the sampling bar», press the manual sampling bar or press the «Startup» key on the front panel of the ABX Micros 60 to start the analysis cycle. The ABX MICROS 60-CS/CT is a fully automated (Microprocessor controlled) Hematology analyzer used for in-vitro diagnostics testing of Whole Blood specimens, Platelet PRP samples, and Whole Blood component concentrates. The ABX MICROS 60-CS/CT is available in 5, 8, 16, and 18 parameters. These parameters are noted according to the system setup. ABX Pentra XL 80; ABX Pentra 60 C+; ABX Micros Series; Yumizen H2500/H1500; Yumizen P8000; Yumizen T6000; Yumizen SPS; QAP; Digital Cell Morphology Systems; Global Hematology Solutions; Global Hemostasis Solutions; Clinical Chemistry.

  • Impedance technology, photometry
  • Microsampling: Only 10 µL
  • No compressor (stepper motor technology)
  • Dedicated infectious disease sample containment for labs

ABX Micros 60 Concept And Technology

  • Measurement methods: Impedance technology, photometry.
  • Liquid valves (without pinching): Reliability and increased precision
  • No compressor (stepper motor): Reliability, less maintenance and noise

Manufactured by HORIBA ABX SAS

Technical

Weight

31 lbs

Power supply

90/240 V - 50/60 Hz - 230 VA.

Dimensions

16.5 in x 14 x 12.5 in (HxWxD).

Printer

Laser Printer

Throughout

60 samples/hour (on open tube)
55 samples/hour (closed tube)

Data management

RS 232 output.
Uni-directional connection.

Memory

LiteDM Patient Data Management System

Screen

2 lines of 40 alphanumerical characters.

Identification

Barcode reader (optional).

Samples

Abx Micros 60 Hematology Analyzer User Manual User

60 samples/hour (open tube)

Parameters

WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, PLT, MPV, GRA%/#, LYM %/# and MON%/# God hand free download for ppsspp.

Graphics

WBC, RBC and PLT histograms

Sample-Volume

CBC mode: 10 µL

Stability

Little snitch photoshop cs6. 24 hours (CBC)

Operating mode

Open tube mode with 4 position tube holder (Cal, 125 or 500 µL, 3 or 5 mL)

Identification

Patient I.D., Sequence#

Flags

Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags

Results

Parameters

Precision (cv %)

Linearity

/virtual-dj-8-evolution-full-download.html. WBC

< 2,5%

0-100 x 10³/mm3

RBC

< 2%

0-8 x 106/mm3

HGB

< 1.5%

0-26 g/dl

HCT

< 2%

0-80%

PLT (whole blood)

< 5%

0-2200 x 10³/mm3

PLT (PLT concentrate)

< 5%

0-4000 x 10³/mm3

Abx Micros 60 Hematology Analyzer User Manuals

ABX Micros 60CS Hematology Analyzer
Impedance technology, photometry
Microsampling- Only 10 µL
Smart card based memory storage (optional)
No compressor (stepper motor technology)
Features:
ABX Micros 60 Concept And Technology
Measurement methods: Impedance technology, photometry.
Liquid valves (without pinching): Reliability and increased precision
No compressor (stepper motor): Reliability, less maintenance and noise
Specifications:
Weight
31 lbs
Power supply
90/240 V - 50/60 Hz - 230 VA.
Dimensions
16.5 in x 14 x 12.5 in (HxWxD).
Throughout
60 samples/hour (on open tube)
55 samples/hour (closed tube)
Data management
Automated smart cards used for QC and patient data storage.
RS 232 output.
Uni-directional connection.
Memory
Last result (with curves) or 78 results (smart card based memory)
Screen
2 lines of 40 alphanumerical characters.
Samples
60 samples/hour (open tube)
Parameters (16)
CBC Mode (16 parameters): WBC, RBC, HGB, HCT, MCV, RDW, MCH, MCHC, PLT, MPV,*PCT,*PDW, % and # of LYM, MON and GRA
Graphics
LMG histogram, RBC, PLT histograms
Sample-Volume
CBC mode : 10 µL
Stability
24 hours (CBC)
Operating mode
Open tube mode with 4 position tube holder (Cal, 125 or 500 µL, 3 or 5 mL)
Identification
Patient I.D., Sequence#
Flags
Pathological flags (WBC, RBC, and PLT) normal limits and counting failure flags
* RUO (Research Use Only)
Results
Parameters
Precision (cv %)
Linearity
WBC
2,5%
0-100 x 10³/mm3
RBC
2%
0-8 x 106/mm3
HGB
1.5%
0-26 g/dl
HCT
2%
0-80%
PLT (whole blood)
5%
0-2200 x 10³/mm3
PLT (PLT concentrate)
5%
0-4000 x 10³/mm3
REFURBISHED ITEM
IT COMES WITH 90 DAYS PARTS WARRANTY
all parts are label and marks to keep the warranty.